PhdDevice

PhdDevice

activePersonal Health Device Implementation Guide2.0.0resource35 key elements

Profile for the Device Resource for a PHD

Metadata

hl7.org/fhir
Canonical URL
http://hl7.org/fhir/uv/phd/StructureDefinition/PhdDevice
ID
PhdDevice
Type
Base Definition
Derivation
constraint
Mandatory and Must-Support Elements

Elements with cardinality > 0 or marked as Must Support (S)

PathCard.TypeFlagsDescription
identifier
1+
Information that uniquely describes the personal health device
identifier.type
1+
Required IEEE 11073-10206 System Id code system coding
identifier.system
1+
EUI-64 system identifier
identifier.value
1+
System id value from System-Id attribute
identifier.type
1+
Required Bluetooth MAC address code system coding
identifier.system
1+
identifier.value
1+
The Bluetooth MAC as an EUI-48 such as 00-E5-DE-AD-77-C3.
identifier.type
1+
Required Ethernet MAC address code system coding
identifier.system
1+
identifier.value
1+
The MAC as an EUI-48 such as 00-E5-DE-AD-77-C8.
udiCarrier.carrierHRF
1+
The UDI Human Readable Barcode String form is supported in ACOM and GHS.
udiCarrier.entryType
1+
manufacturer
1+
Manufacturer name from SystemInfo.system-manufacturer
modelNumber
1+
Model number from SystemInfo.model-number
type
1+
type.coding
1+
:PhdCode1..1
1+
Indicates the device is a PHD.
specialization
1+
:MDCType1..*
1+
This element contains an entry for each supported specialization
specialization.systemType.coding
1+
specialization.version
1+
The version of the specialization standard supported by the PHD from the System-Type-Spec-List[i].version
version
1+
A PHD may report firmware, hardware, software, internal protocol, nomenclature and ACOM versions.
:MDCType1..*
1+
Required MDC device version type entry
property.valueCode.coding
1+
property.valueCode1..1
1+
property.valueCode.coding.system
1+
property.valueCode.coding.code
1+
If bit is set, code contains Y if cleared, N
property.valueCode1..1
1+
property.valueCode.coding.system
1+
property.valueCode.coding.code
1+
If bit is set, the device is not regulated. If cleared, the device is regulated.
property.valueQuantity1..1
1+
The value. All the time fields are scaled to microseconds
property.valueQuantity.system
1+
property.valueQuantity1..1
1+
The value. All the time fields are scaled to microseconds
property.valueQuantity.system
1+
property.valueCode.text
1+
The USB Vendor and Product ID as reported by the PHD in the form of a coded string such as '1234:56AB'.