Safety

Safety

draftPoint-of-Care Devices0.3.0complete4 concepts

CodeSystem for IEEE 11073-10207 SafetyClassification

Metadata

hl7.org/fhir
Canonical URL
http://hl7.org/fhir/uv/pocd/CodeSystem/safety
ID
safety
Version
0.3.0
Publisher
HL7 International - Devices Work Group
Case Sensitive
No

Concepts (4)

CodeDisplayDefinition
InfInformationalThe descriptor and the related state information are intended to be used for information purposes only. They are not intended to be used in clinical functions.
MedAMedical Class AMedical Class A. The descriptor and related state information are intended to be used in clinical functions, specifically for general display in order to support patient and device monitoring. The displayed data is not intended to be used as sole source for diagnostic or therapeutic decisions. Deviations from this intended use are in the sole responsibility of the SERVICE CONSUMER.
MedBMedical Class BAThe descriptor and related state information are intended to be used in clinical functions. The manufacturer has specified and considered a specific intended use for the data, which could result in nonserious injury. Deviations from this intended use are in the sole responsibility of the SERVICE CONSUMER.
MedCMedical Class CThe descriptor and related state information are intended to be used in clinical functions. The manufacturer has specified and considered a specific intended use for the data, which could result in serious injury. Deviations from this intended use are in the sole responsibility of the SERVICE CONSUMER.