OUL_R24

OUL^R24 Acknowledgment Choreography

OUL^R24
This message was designed to accommodate multi-specimen oriented testing. It should be applicable to, e.g., laboratory automation systems requiring container.

Generally this construct allows transfer of multiple results, each one related to none, one or more specific containers with one or more specimens from a patient. (Example: Creatinine Clearance result with detailed information about the urine and serum specimens and their containers.)

Acknowledgment Rules(4)

ModeMSH-15MSH-16Immediate ACKApplication ACK
Originalblankblank-
^^
EnhancedNENE--
EnhancedNEAL,SU,ER-
^^
EnhancedAL,SU,ERAL,SU,ER
^^
^^

Legend

blank - Field is empty/not valued
NE - Never (no acknowledgment)
AL - Always
SU - Success only
ER - Error/reject only