DICOM Module

Breast Tomosynthesis Acquisition Module

Attributes
42
Required (Type 1 / 1C)
24
Source
PS3.3 §C.8.21.3.4

Overview

specifies the Attributes of the .

Attributes

Type: 1 required (non-zero), 1C required when condition met, 2 required (may be zero-length), 2C conditional, 3 optional.

Attribute Tag Type
X-Ray 3D Acquisition Sequence (0018,9507) 1
Field of View Shape (0018,1147) 1
X-Ray Receptor Type (0018,9420) 1
Include >Include
Include >Include
Distance Source to Detector (0018,1110) 1
Distance Source to Patient (0018,1111) 1
Estimated Radiographic Magnification Factor (0018,1114) 1
Anode Target Material (0018,1191) 1
Body Part Thickness (0018,11A0) 1
Exposure Control Mode (0018,7060) 1
Exposure Control Mode Description (0018,7062) 1
Half Value Layer (0040,0314) 1
Organ Dose (0040,0316) 3
Entrance Dose in mGy (0040,8302) 3
Entrance Dose Derivation (0040,8303) 3
Focal Spot(s) (0018,1190) 1
Detector Binning (0018,701A) 1C
Detector Temperature (0018,7001) 1
Filter Type (0018,1160) 1
Filter Material (0018,7050) 1
Filter Thickness Minimum (0018,7052) 3
Filter Thickness Maximum (0018,7054) 3
Filter Beam Path Length Minimum (0018,7056) 3
Filter Beam Path Length Maximum (0018,7058) 3
Compression Force (0018,11A2) 1
Compression Pressure (0018,11A3) 3
Compression Contact Area (0018,11A5) 3
Paddle Description (0018,11A4) 1
Per Projection Acquisition Sequence (0018,9538) 1
Include >>Include
Positioner Primary Angle (0018,1510) 1
Positioner Primary Angle Direction (0018,9559) 3
Positioner Secondary Angle (0018,1511) 1C
Exposure Time in ms (0018,9328) 1
Exposure in mAs (0018,9332) 1
Relative X-Ray Exposure (0018,1405) 1
Organ Dose (0040,0316) 3
Entrance Dose in mGy (0040,8302) 3
Entrance Dose Derivation (0040,8303) 3
Include >>Include
Irradiation Event UID (0008,3010) 3

Used in 1 CIOD