DICOM Module

MR Image Module

Attributes
53
Required (Type 1 / 1C)
8
Source
PS3.3 §C.8.3.1

Overview

specifies the Attributes of the , which describe MR images.

Attributes

Type: 1 required (non-zero), 1C required when condition met, 2 required (may be zero-length), 2C conditional, 3 optional.

Attribute Tag Type
Image Type (0008,0008) 1
Samples per Pixel (0028,0002) 1
Photometric Interpretation (0028,0004) 1
Bits Allocated (0028,0100) 1
Bits Stored (0028,0101) 1
High Bit (0028,0102) 1
Scanning Sequence (0018,0020) 1
Sequence Variant (0018,0021) 1
Scan Options (0018,0022) 2
MR Acquisition Type (0018,0023) 2
Repetition Time (0018,0080) 2C
Echo Time (0018,0081) 2
Echo Train Length (0018,0091) 2
Inversion Time (0018,0082) 2C
Trigger Time (0018,1060) 2C
Sequence Name (0018,0024) 3
Angio Flag (0018,0025) 3
Number of Averages (0018,0083) 3
Imaging Frequency (0018,0084) 3
Imaged Nucleus (0018,0085) 3
Echo Number(s) (0018,0086) 3
Magnetic Field Strength (0018,0087) 3
Number of Phase Encoding Steps (0018,0089) 3
Percent Sampling (0018,0093) 3
Percent Phase Field of View (0018,0094) 3
Pixel Bandwidth (0018,0095) 3
Nominal Interval (0018,1062) 3
Beat Rejection Flag (0018,1080) 3
Low R-R Value (0018,1081) 3
High R-R Value (0018,1082) 3
Intervals Acquired (0018,1083) 3
Intervals Rejected (0018,1084) 3
PVC Rejection (0018,1085) 3
Skip Beats (0018,1086) 3
Heart Rate (0018,1088) 3
Cardiac Number of Images (0018,1090) 3
Trigger Window (0018,1094) 3
Reconstruction Diameter (0018,1100) 3
Receive Coil Name (0018,1250) 3
Transmit Coil Name (0018,1251) 3
Acquisition Matrix (0018,1310) 3
In-plane Phase Encoding Direction (0018,1312) 3
Flip Angle (0018,1314) 3
SAR (0018,1316) 3
Variable Flip Angle Flag (0018,1315) 3
dB/dt (0018,1318) 3
Temporal Position Identifier (0020,0100) 3
Number of Temporal Positions (0020,0105) 3
Temporal Resolution (0020,0110) 3
Include Include
Include Include
Include Include
B1rms (0018,1320) 3

Used in 1 CIOD