DICOM Module

NM Image Module

Attributes
19
Required (Type 1 / 1C)
3
Source
PS3.3 §C.8.4.9

Overview

specifies the Attributes of the , which describe Nuclear Medicine Images.

Attributes

Type: 1 required (non-zero), 1C required when condition met, 2 required (may be zero-length), 2C conditional, 3 optional.

Attribute Tag Type
Image Type (0008,0008) 1
Image ID (0054,0400) 3
Lossy Image Compression (0028,2110) 1C
Counts Accumulated (0018,0070) 2
Acquisition Termination Condition (0018,0071) 3
Table Height (0018,1130) 3
Table Traverse (0018,1131) 3
Actual Frame Duration (0018,1242) 1C
Count Rate (0018,1243) 3
Processing Function (0018,5020) 3
Corrected Image (0028,0051) 3
Whole Body Technique (0018,1301) 3
Scan Velocity (0018,1300) 2C
Scan Length (0018,1302) 2C
Trigger Source or Type (0018,1061) 3
Include Include
Real World Value Mapping Sequence (0040,9096) 3
Include >Include
Scan Progression Direction (0054,0501) 3

Used in 1 CIOD