DICOM Module
NM Phase Module
- Attributes
- 8
- Required (Type 1 / 1C)
- 5
- Source
- PS3.3 §C.8.4.14
Overview
specifies the Attributes of the , which describe dynamic phases of a dynamic acquisition image performed on the patient. This Module is present only when Image Type (0008,0008), Value 3, is equal to DYNAMIC. A phase is defined as a collection of Frames in which the acquisition time per Frame and the time delay between Frames remains constant. A new phase shall be defined whenever there is a change in the time between Frames, the acquisition time per Frame, or the position of the patient relative…
Attributes
Type: 1 required (non-zero), 1C required when condition met, 2 required (may be zero-length), 2C conditional, 3 optional.
| Attribute | Tag | Type |
|---|---|---|
| Phase Information Sequence | (0054,0032) | 2C |
| Phase Delay | (0054,0036) | 1 |
| Actual Frame Duration | (0018,1242) | 1 |
| Pause Between Frames | (0054,0038) | 1 |
| Number of Frames in Phase | (0054,0033) | 1 |
| Trigger Vector | (0054,0210) | 3 |
| Number of Triggers in Phase | (0054,0211) | 1C |
| Phase Description | (0054,0039) | 3 |