DICOM Module

Photoacoustic Acquisition Parameters Module

Attributes
9
Required (Type 1 / 1C)
3
Source
PS3.3 §C.8.34.2

Overview

This section describes Photoacoustic Imaging parameters that are constant for the acquisition of all Frames in this image.

Attributes

Type: 1 required (non-zero), 1C required when condition met, 2 required (may be zero-length), 2C conditional, 3 optional.

Attribute Tag Type
Excitation Wavelength Sequence (0018,9825) 1
Excitation Wavelength (0018,9826) 1
Illumination Translation Flag (0018,9828) 3
Illumination Type Code Sequence (0022,0016) 3
Include >Include
Acoustic Coupling Medium Flag (0018,9829) 1
Acoustic Coupling Medium Code Sequence (0018,982A) 2C
Include >Include
Acoustic Coupling Medium Temperature (0018,982B) 3

Used in 1 CIOD