DICOM Module

RT Brachy Application Setups Module

Attributes
98
Required (Type 1 / 1C)
36
Source
PS3.3 §C.8.8.15

Overview

The describes the application of a brachytherapy radiotherapy treatment. It contains one or more sources, each associated with one or more Channels. A Channel is a device by which a source is placed in its intended treatment position or positions. A Channel may consist of a Source Applicator plus a Transfer Tube, a Source Applicator alone, a rigid or flexible linear source, or a seed. A number of Channels (for example applicators, sources or seeds) are generally arranged in an Application Setup…

Attributes

Type: 1 required (non-zero), 1C required when condition met, 2 required (may be zero-length), 2C conditional, 3 optional.

Attribute Tag Type
Brachy Treatment Technique (300A,0200) 1
Brachy Treatment Type (300A,0202) 1
Treatment Machine Sequence (300A,0206) 1
Treatment Machine Name (300A,00B2) 2
Manufacturer (0008,0070) 3
Institution Name (0008,0080) 3
Institution Address (0008,0081) 3
Institutional Department Name (0008,1040) 3
Institutional Department Type Code Sequence (0008,1041) 3
Include >>Include
Manufacturer's Model Name (0008,1090) 3
Device Serial Number (0018,1000) 3
Date of Manufacture (0018,1204) 3
Date of Installation (0018,1205) 3
Source Sequence (300A,0210) 1
Source Number (300A,0212) 1
Source Serial Number (3008,0105) 3
Source Model ID (300A,021B) 3
Source Description (300A,021C) 3
Source Type (300A,0214) 1
Source Manufacturer (300A,0216) 3
Active Source Diameter (300A,0218) 3
Active Source Length (300A,021A) 3
Material ID (300A,00E1) 3
Source Encapsulation Nominal Thickness (300A,0222) 3
Source Encapsulation Nominal Transmission (300A,0224) 3
Source Isotope Name (300A,0226) 1
Source Isotope Half Life (300A,0228) 1
Source Strength Units (300A,0229) 1C
Reference Air Kerma Rate (300A,022A) 1
Source Strength (300A,022B) 1C
Source Strength Reference Date (300A,022C) 1
Source Strength Reference Time (300A,022E) 1
Application Setup Sequence (300A,0230) 1
Application Setup Type (300A,0232) 1
Application Setup Number (300A,0234) 1
Application Setup Name (300A,0236) 3
Application Setup Manufacturer (300A,0238) 3
Template Number (300A,0240) 3
Template Type (300A,0242) 3
Template Name (300A,0244) 3
Referenced Reference Image Sequence (300C,0042) 3
Include >>Include
Total Reference Air Kerma (300A,0250) 1
Brachy Accessory Device Sequence (300A,0260) 3
Brachy Accessory Device Number (300A,0262) 2
Brachy Accessory Device ID (300A,0263) 2
Brachy Accessory Device Type (300A,0264) 1
Brachy Accessory Device Name (300A,0266) 3
Material ID (300A,00E1) 3
Brachy Accessory Device Nominal Thickness (300A,026A) 3
Brachy Accessory Device Nominal Transmission (300A,026C) 3
Referenced ROI Number (3006,0084) 2
Channel Sequence (300A,0280) 1
Channel Number (300A,0282) 1
Channel Length (300A,0284) 2
Channel Total Time (300A,0286) 1
Source Movement Type (300A,0288) 1
Number of Pulses (300A,028A) 1C
Pulse Repetition Interval (300A,028C) 1C
Source Applicator Number (300A,0290) 3
Source Applicator ID (300A,0291) 2C
Source Applicator Type (300A,0292) 1C
Source Applicator Name (300A,0294) 3
Source Applicator Length (300A,0296) 1C
Source Applicator Tip Length (300A,0274) 2C
Source Applicator Manufacturer (300A,0298) 3
Material ID (300A,00E1) 3
Source Applicator Wall Nominal Thickness (300A,029C) 3
Source Applicator Wall Nominal Transmission (300A,029E) 3
Source Applicator Step Size (300A,02A0) 1C
Applicator Shape Referenced ROI Number (300A,02A1) 3
Referenced ROI Number (3006,0084) 2C
Transfer Tube Number (300A,02A2) 2
Transfer Tube Length (300A,02A4) 2C
Channel Effective Length (300A,0271) 3
Channel Inner Length (300A,0272) 2C
Afterloader Channel ID (300A,0273) 3
Channel Shield Sequence (300A,02B0) 3
Channel Shield Number (300A,02B2) 1
Channel Shield ID (300A,02B3) 2
Channel Shield Name (300A,02B4) 3
Material ID (300A,00E1) 3
Channel Shield Nominal Thickness (300A,02B8) 3
Channel Shield Nominal Transmission (300A,02BA) 3
Referenced ROI Number (3006,0084) 2
Referenced Source Number (300C,000E) 1
Number of Control Points (300A,0110) 1
Final Cumulative Time Weight (300A,02C8) 1C
Brachy Control Point Sequence (300A,02D0) 1
Control Point Index (300A,0112) 1
Cumulative Time Weight (300A,02D6) 2
Control Point Relative Position (300A,02D2) 1
Control Point 3D Position (300A,02D4) 3
Control Point Orientation (300A,0412) 3
Brachy Referenced Dose Reference Sequence (300C,0055) 3
Referenced Dose Reference Number (300C,0051) 1
Cumulative Dose Reference Coefficient (300A,010C) 1

Used in 1 CIOD