DICOM Module

RT Dose Contribution Record Module

Attributes
22
Required (Type 1 / 1C)
15
Source
PS3.3 §C.36.21

Overview

specifies the Attributes of the , which describe the delivered dose.

Attributes

Type: 1 required (non-zero), 1C required when condition met, 2 required (may be zero-length), 2C conditional, 3 optional.

Attribute Tag Type
Radiation Dose Identification Sequence (300A,0618) 1
Radiation Dose Identification Index (300A,0603) 1
Referenced Radiation Dose Identification Index (300A,060C) 1C
Radiation Dose Identification Label (300A,0619) 1
Conceptual Volume Sequence (3010,0025) 1
Include >>Include
Radiation Dose Sequence (300A,0617) 1
Referenced RT Radiation Record Sequence (300A,0703) 1
Include >>Include
Radiation Dose Values Parameters Sequence (300A,061F) 1C
Referenced Radiation Dose Identification Index (300A,060C) 1
Include >>Include
Meterset to Dose Mapping Sequence (300A,0620) 1
Cumulative Meterset (300A,063C) 1
Radiation Dose Value (300A,0625) 1
Measured Meterset to Dose Mapping Sequence (300A,0772) 1C
Referenced Expected In-Vivo Measurement Value Index (300A,0773) 3
Cumulative Meterset (300A,063C) 1
Radiation Dose Value (300A,0625) 1
Dose Measurement Device Code Sequence (300A,0774) 2
Include >>>Include
Measured Dose Description (3008,0012) 3

Used in 1 CIOD