Overview
The RT Ion Tolerance Tables Module contains information describing the maximum allowed differences between the planned and measured Attributes for Ion therapy.
Attributes
Type: 1 required (non-zero), 1C required when condition met, 2 required (may be zero-length), 2C conditional, 3 optional.
| Attribute | Tag | Type |
|---|---|---|
| Ion Tolerance Table Sequence | (300A,03A0) | 1 |
| Tolerance Table Number | (300A,0042) | 1 |
| Tolerance Table Label | (300A,0043) | 3 |
| Gantry Angle Tolerance | (300A,0044) | 3 |
| Beam Limiting Device Angle Tolerance | (300A,0046) | 3 |
| Beam Limiting Device Tolerance Sequence | (300A,0048) | 3 |
| RT Beam Limiting Device Type | (300A,00B8) | 1 |
| Beam Limiting Device Position Tolerance | (300A,004A) | 1 |
| Patient Support Angle Tolerance | (300A,004C) | 3 |
| Table Top Vertical Position Tolerance | (300A,0051) | 3 |
| Table Top Longitudinal Position Tolerance | (300A,0052) | 3 |
| Table Top Lateral Position Tolerance | (300A,0053) | 3 |
| Table Top Pitch Angle Tolerance | (300A,004F) | 3 |
| Table Top Roll Angle Tolerance | (300A,0050) | 3 |
| Snout Position Tolerance | (300A,004B) | 3 |
| Head Fixation Angle Tolerance | (300A,0152) | 3 |
| Chair Head Frame Position Tolerance | (300A,0153) | 3 |
| Fixation Light Azimuthal Angle Tolerance | (300A,0154) | 3 |
| Fixation Light Polar Angle Tolerance | (300A,0155) | 3 |