DICOM Module
Waveform Annotation Module
- Attributes
- 19
- Required (Type 1 / 1C)
- 10
- Source
- PS3.3 §C.10.10
Overview
The table in this section contains Attributes that identify annotations to the waveform of the current SOP Instance. Each annotation conceptually forms the equivalent of a overlay on a presentation display of the annotated entity. Annotations may represent a measurement or categorization based on the waveform data, identification of regions of interest or particular features of the waveform, or events during the data collection that may affect diagnostic interpretation (e.g., the time at which…
Attributes
Type: 1 required (non-zero), 1C required when condition met, 2 required (may be zero-length), 2C conditional, 3 optional.
| Attribute | Tag | Type |
|---|---|---|
| Waveform Annotation Sequence | (0040,B020) | 1 |
| Unformatted Text Value | (0070,0006) | 1C |
| Concept Name Code Sequence | (0040,A043) | 1C |
| Include | >>Include | — |
| Modifier Code Sequence | (0040,A195) | 1C |
| Include | >>>Include | — |
| Concept Code Sequence | (0040,A168) | 3 |
| Include | >>Include | — |
| Modifier Code Sequence | (0040,A195) | 1C |
| Include | >>>Include | — |
| Numeric Value | (0040,A30A) | 3 |
| Measurement Units Code Sequence | (0040,08EA) | 3 |
| Include | >>Include | — |
| Referenced Waveform Channels | (0040,A0B0) | 1 |
| Temporal Range Type | (0040,A130) | 1C |
| Referenced Sample Positions | (0040,A132) | 1C |
| Referenced Time Offsets | (0040,A138) | 1C |
| Referenced DateTime | (0040,A13A) | 1C |
| Annotation Group Number | (0040,A180) | 3 |
Used in 16 CIODs
- 12-Lead ECG —
- Ambulatory ECG —
- Arterial Pulse Waveform —
- Basic Cardiac Electrophysiology Waveform —
- Basic Voice Audio Waveform U
- Body Position Waveform —
- Electromyogram —
- Electrooculogram —
- General 32-bit ECG —
- General Audio Waveform —
- General ECG —
- Hemodynamic Waveform —
- Multi-channel Respiratory Waveform —
- Respiratory Waveform —
- Routine Scalp Electroencephalogram —
- Sleep Electroencephalogram —